P447 – Beam Shaping Slits I14 DLS
Application
Beam shaping slits for hard X-ray synchrotron radiation
Year of delivery
2016
Installation site
Beamline I14 at Diamond Light Source, Didcot, United Kingdom
Design Features
-
Double slit mechanism – vertically and horizontally defining slits.
-
Motorized and encoded slits opening, symmetric opening, non-linear opening characteristics.
- Tungsten slit blades.
- Electrical isolation of all slit blades.
- Isolated bias plate.
- Adjustable support structure.
Performance Features
- Precision and slit opening repeatability smaller than 100 nm.
- Parallelism smaller than 1 µm over 10 mm edge length.
- Straightness smaller than 0.3 µm over 10 mm edgle length.
- Very small deviation (submicrometer) of the slit opening calibration curve from the theoretical non-linear slit opening function.
Outer Dimensions
Technical specifications and performance values
Chamber type
cylindrical, horizontal axis
Main flanges
DN160CF cluster flanges
Material
1.4301
Base pressure
10-10 mbar range
Material
Tungsten
Straightness
< 0.25 µm over 10 mm (1σ)
Parallelism
-0.09 µm /mm
(15 µm slit opening)
-0.004 µm /mm
(95 µm slit opening)
Range
5 µm to 8.5 mm
Repeatability
± 0.04 µm
(up to 435 µm slit opening)
Resolution (design value)
0.006 µm /motor fine step
(1/16, recommended)
0.05 µm /encoder count
Material
Tungsten
Straightness
< 0.15 µm over 10 mm (1σ)
Parallelism
0.02 µm /mm
(15 µm slit opening)
-0.008 µm /mm
(95 µm slit opening)
Range
5 µm to 10 mm
Repeatability
± 0.038 µm
(up to 350 µm slit opening)
Resolution (design value)
0.006 µm /motor fine step
(1/16, recommended)
0.05 µm /encoder count