P486 - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
P486 - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain

P486 – Exit Slit Unit XAIRA ALBA CELLS

Application

Exit slit unit for soft X-ray radiation

Year of delivery

2021

Installation site

XAIRA beamline at ALBA of CELLS, Spain

Design Features

  • Double slit mechanism – vertically and horizontally defining slits.
  • Motorized and encoded slits opening, symmetric opening, non-linear opening characteristics.
  • Motorized and encoded vertical translation of the vertically defining slit.
  • Tungsten slit blades.

  • Motorized and encoded slit unit horizonatal transversal translation.

  • Adjustable support structure.

Performance Features

  • Precision and slit opening repeatability smaller than 150 nm.
  • Parallelism smaller than 2 µm over 10 mm edge length.
  • Straightness smaller than 2 µm over 10 mm edge length.
  • Small deviation of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P486 - Outer dimensions - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Horizontal transversal translation

Range

-9.54 mm to 8.12 mm

Repeatability

0.40 µm

Hysteresis

4 µm

Linearity

1.90 µm

Parasitic roll

11.60 µrad

Resolution (design value)

0.003 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.10 µm

Vertically defining slit (VDS)

Slit blades

Material

Tungsten

Straightness

< 1 µm over 10 mm (1σ)

Parallelism

-0.02 µm /mm at 27 µm slit opening
0.02 µm /mm at 60 µm slit opening
-0.06 µm /mm at 110 µm slit opening

Slit opening

Range

3 µm to 4 mm

Repeatability

0.05 µm at 27 µm slit opening

0.09 µm at 60 µm slit opening

0.07 µm at 110 µm slit opening

Slit center stability (> 1h)

0.026 µm

Resolution (design value)

0.00156 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.50 µm

Slit vertical translation

Range

-2.34 µm to 2.31 mm

Repeatability

0.20 µm from 2 mm to -0.15 mm
0.06 µm from 2 mm to -0.10 mm
0.08 µm from 2 mm to 0 mm
0.15 µm fom 2 mm to 0.10 mm

Hysteresis

2 µm

Linearity

0.22 µm

Resolution (design value)

0.00156 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.50 µm

Horizontally defining slit (HDS)

Slit blades

Material

Tungsten

Straightness

< 2 µm over 10 mm (1σ)

Parallelism

-0.01 µm /mm at 23 µm slit opening
-0.16 µm /mm at 50 µm slit opening
-0.04 µm /mm at 100 µm slit opening

Slit opening

Range

2.90 µm to 5 mm

Repeatability

0.10 µm at 24 µm slit opening

0.07 µm at 58 µm slit opening

0.05 µm at 105 µm slit opening

Slit center stability (2h)

0.038 µm

Resolution (design value)

0.00156 µm /motor fine step
(1/64, recommended)
0.05 µm /encoder count

Resolution (measured value)

0.50 µm

Performance Test Results

Slit unit translation linearity
P486 - Slit unit translation linearity - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit unit translation hysteresis
P486 - Slit unit translation hysteresis - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit unit translation repeatability
P486 - Slit unit translation repeatability - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Parasitic roll after slit unit translation
P486 - Parasitic roll after slit unit translation - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit unit translation positioning resolution
P486 - Slit unit translation positioning resolution - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit unit translation positioning resolution
P486 - Slit unit translation positioning resolution - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit width calibration (VDS)
P486 - Slit width calibration (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit opening repeatability (VDS)
P486 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit opening positioning resolution (VDS)
P486 - Slit opening positioning resolution (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit center stability (VDS)
P486 - Slit center stability (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit blades parallelism (VDS)
P486 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit blades parallelism (VDS)
P486 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit blades parallelism (VDS)
P486 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit translation linearity (VDS)
P486 - Slit translation linearity (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit translation hysteresis (VDS)
P486 - Slit translation hysteresis (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit translation repeatability (VDS)
P486 - Slit translation repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit translation positioning resolution (VDS)
P486 - Slit translation positioning resolution - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit width calibration (HDS)
P486 - Slit width calibration (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit opening repeatability (HDS)
P486 - Slit opening repeatability (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit opening positioning resolution (HDS)
P486 - Slit opening positioning resolution (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit center stability (HDS)
P486 - Slit center stability (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit blades parallelism (HDS)
P486 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit blades parallelism (HDS)
P486 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Slit blades parallelism (HDS)
P486 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain
Residual gas analysis
P486 - Residual gas analysis - Exit slit unit for soft X-ray synchrotron radiation, BL06-XAIRA beamline at ALBA of CELLS, Spain